medical MODERATE Updated 2026-08-26

Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI), Automated Field Steam Steri

See all recalls from Fort Defiance Industries, LLC →

Recalled Product

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Code: All serial numbers/ GTIN: (01) 00862992000327

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
View official government recall