Medline Industries, LP recalls Medline breathing circuits labeled as: 1) HUDSON RCI, CIR

See all recalls from Medline Industries, LP →

Recalled Product

Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD7801780-18) (Teleflex Product Number/SKU 780-18).

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 1) Medline Product Number/SKU HUD78007, UDI/DI 10197344042463 (each) 20197344042460 (case), All lot numbers; 2) Medline Product Number/SKU HUD78009, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 3) Medline Product Number/SKU 780-13, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 4) Medline Product Number/SKU 780-14, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 5) Medline Product Number/SKU 780-15, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 6) Medline Product Number/SKU HUD78018, UDI/DI 10197344010660 (each) 20197344010667 (case), All lot numbers.

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
View official government recall